CAPILLARYS 3 DBS

The high throughput solution for screening newborn hemoglobin disorders.

Sebia_courbe_1600x200_

About CAPILLARYS 3 DBS

Capillary electrophoresis is a proven technology(1),(2),(3) screening for newborn hemoglobin disorders and provides high quality results.
Sebia CAPILLARYS 3 DBS is the latest generation of instrument dedicated to newborn hemoglobin screening, offering a smooth workflow and excellent analytical features.

The CAPILLARYS 3 DBS instrument is an automated, multitasking capillary electrophoresis instrument using 12 capillaries simultaneously for hands-free electrophoretic separation at high throughput.

The CAPILLARYS 3 DBS enables the analysis of samples from microplate wells and provides fully automated electrophoresis separation, from the dried blood spots on filter paper (Guthrie card) through to the final electrophoretic pattern: sample identification, sample dilution, sample incubation, capillary washing, sample injection into the capillaries, migration, detection, processing of the results, and data transmission to LIS.

This instrument is an innovative alternative to commonly used technology such as IEF and HPLC for all the laboratories involved in newborn hemoglobin screening, bringing seamless sample traceability, high quality results, and high throughput while offering a reduced hands-on time.

Key benefits

High throughput instrument

Up to 70 samples/hour, enabling Newborn Screening laboratories to efficiently manage their workload and deliver results to pathologists and clinicians quickly.

Full traceability

From sample punch to result, there is no interruption in the traceability chain, preventing transcription mistakes (e.g. errors from manually entering the sample ID into the analysis software).

Walkaway Automation

Load up to 8, 96-well microplates with continuous loading. This saves hands on time and lets the lab staff focus on important tasks.

Automatic Reagent Control

RFID tags for all reagents and Hb AF Control ensure a smooth and efficient reagent management, reducing user intervention.

Up to
70 results/hour
Up to
70 results/hour

Main Characteristics

Test Menu
  • Hb Neonat technique

 

Sample types: dried blood spot (3.2mm)

Performance

With 12 capillaries in parallel, the CAPILLARYS 3 DBS can process a large number of samples and can analyse up to eight 96-well microplates without reagent refill.

 

                          Throughput (tests/hour)

CAPI 3 NEONAT Hb                                          70

Reagent and consumables
  • Flexible positions for reagents allow for the optimization of on-board reagent management
  • RFID identification for all reagents (reagent type, lot number, expiration date, volume) allows comprehensive traceability throughout workflow
Ergonomics
  • Embedded user interface with a touch screen
  • Automatic start-up and shutdown maintenance
  • Visual alarms (with instrument notifications)
  • A dedicated workstation pilots the instrument and manages results
Instrument details
  • Uses barcoded microplates (supplied with the CAPI 3 NEONAT Hb reagent kit – P/N:2506)
  • Directly samples from the elution of dried blood spots on filter paper.
  • Validated with automated PerkinElmer™ puncher instruments Panthera 9 and DBS models.
  • Barcoded microplates enable full sample traceability: from sample card ID through to final result
  • CAPILLARYS 3 DBS – P/N: 1244